Select a known measurement standard, such as a ruler, to be used for calibration. Take a picture of the selected measurement calibration standard.
Now open the picture of the measurement calibration standard.
Determine the magnification value by reading from the DinoLite adjustment dial. Enter this value in the“Input Mag” box located in upper right hand corner of the screen.
Now click on the measurement icon at top of page and Select“Calibration”.
The software will then prompt for taking a measurement on the selected calibration standard.
Take a sample measurement of the selected calibration standard by clicking beginning and ending points across a known measurement value (such as 5mm’s, half inch,etc). A measurement value before calibration will be displayed (5.179mm in the example below)
Enter the exact known value of the measurement sample (5.0mm in this example) in the“Adjusted Mag.” window and then click “OK”.
The software will automatically calibrate the magnification rate and will show the adjusted magnification rate at the upperright corner of the picture.
Now the software is ready to take calibrated measurements with adjusted magnification rate. To verify calibration simply take another measurement sample of a known measurement value.